Abstract

PurposeTo compare the difference and agreement of axial length (AL) and anterior segment parameters obtained from the MYAH device with Pentacam AXL and IOLMaster 700 in myopic children.MethodsThe study included 60 eyes of 60 myopic children. AL, keratometry (K), and horizontal corneal diameter (CD) were measured with Pentacam AXL, IOLMaster 700, and MYAH, respectively. The Friedman test was used to assess the differences. The Intraclass Correlation Coefficient (ICC) and Bland–Altman plots were used to assess the consistency of measurements.ResultsThe mean age was 10.2 ± 1.8 years (7–16 years). No statistically significant difference was determined between the Pentacam AXL, IOLMaster 700, and MYAH devices in terms of mean AL values (23.61 ± 1.42, 23.62 ± 1.45, 23.61 ± 1.42, respectively) (p = 0.06). The difference between devices in the mean steep K, flat K, mean K, and CD was statistically significant but clinically insignificant (steep K; 44.45 ± 1.25, 44.59 ± 1.23, 44.51 ± 1.24, flat K; 43.29 ± 1.28, 43.43 ± 1.29, 43.35 ± 1.30, mean K; 43.85 ± 1.21, 44.00 ± 1.19, 43.94 ± 1.20, and CD; 11.90 ± 0.34, 12.11 ± 0.38, 11.96 ± 0.31, respectively; p < 0.05). ICC and Bland–Altman plot analysis revealed a high correlation between the three devices in AL, steep K, flat K, mean K, and CD measurements.ConclusionThere was a quite good agreement between the MYAH, Pentacam AXL, and IOLMaster 700 devices regarding AL and anterior segment parameters. MYAH provides reliable measurements and will be a good option in the diagnosis of and follow-up with myopic children.