Abstract
ABSTRACT Purpose To compare the inter-device agreement of anterior chamber depth (ACD), central corneal thickness (CCT), flat keratometry (K1), steep keratometry (K2), corneal astigmatism (ΔK), and white-to-white (WTW) measurements. Methods This is a retrospective study with 73 myopic patients (142 eyes). We extracted and compared anterior segment measurements (ACD, CCT, K1, K2, ΔK, and WTW) of four devices (OPD-Scan III, Pentacam HR, Lenstar LS 900, and Galilei G4), and performed pairwise agreement comparisons between them. Results Agreement analyses revealed that the most agreement occurred: in Pentacam HR and Galilei G4 for ACD measurements, in Pentacam HR and Lenstar LS 900 for CCT measurements, in OPD-Scan III and Galilei G4 for WTW measurements, in OPD-Scan III and Pentacam HR for K1 and K2 measurements, and OPD-Scan III and Galilei G4 for ΔK measurements. Conclusion OPD-Scan III is interchangeable with both Galilei G4 and Lenstar LS 900 for WTW measurements. OPD-Scan III and the Pentacam-HR are interchangeable in K1, K2, and ΔK measurements. OPD-Scan III and Lenstar LS 900, and OPD-Scan III and Galilei G4 are interchangeable in K1 and ΔK measurements, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.