Abstract

Intravascular bubble load after decompression can be detected and scored using ultrasound techniques that measure venous gas emboli (VGE). The aim of this study was to analyse the agreement between ultrasonic bubble grades from a handheld self-positioning product, the O'Dive™, and cardiac 2D ultrasound after decompression. VGE were graded with both bilateral subclavian vein Doppler ultrasound (modified Spencer scale) and 2D cardiac images (Eftedal Brubakk scale). Agreement was analysed using weighted kappa (Kw). Analysis with Kw was made for all paired grades, including measurements with and without zero grades, and for each method's highest grades after each dive. A total of 152 dives yielded 1,113 paired measurements. The Kw agreement between ultrasound VGE grades produced by cardiac 2D images and those from the O'Dive was 'fair'; when zero grades were excluded the agreement was 'poor'. The O'Dive was found to have a lower sensitivity to detect VGE compared to 2D cardiac image scoring. Compared to 2D cardiac image ultrasound, the O'Dive yielded generally lower VGE grades, which resulted in a low level of agreement (fair to poor) with Kw.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.