Abstract

The purpose of this study was to evaluate the agreement between three methods for measuring pupil size in patients implanted with multifocal intraocular lenses (MIOLs): Keratograph 5M (K5M), Pentacam AXL Wave (PW), and a simple hand ruler. Sixty-nine subjects implanted with MIOLs and measured at the three-month follow-up visit were included in this retrospective analysis. K5M and PW were used to measure the photopic (PP) and mesopic (MP) pupil sizes, and a hand ruler was used to measure the pupil under environmental light conditions (135 lux). The Bland-Altman method with its limits (LoAs) was used to assess the agreement. The median PP was 2.8, 2.95, and 3 mm for K5M, PW, and the ruler, respectively (p < 0.05). Differences in PP were statistically significant for all paired comparisons (p < 0.0005) except between PW and the ruler (p = 0.44). The LoAs for the difference in PP between K5M and PW was 0.63 mm. The mean difference for MP between K5M and PW was 0.04 mm (p = 0.34) with LoAs of 0.72 mm. MP measured with K5M and PW could be considered interchangeable, although a correction of -0.3 mm (IC95%: -0.23 to -0.39) should be applied to PP measured with PW to attain the K5M mean.