Abstract

The purpose of this analysis was to study agreement and trending of stroke volume measured by oesophageal Doppler and 3rd generation Vigileo during fluid loads in patients with severe aortic stenosis. Observational study in 32 patients (30 analyzed) scheduled for aortic valve replacement due to severe aortic stenosis. After induction of anesthesia and before start of surgery, hemodynamic registrations for 1 min were obtained before and after a fluid load. Agreement between stroke volume measured by oesophageal Doppler (SVOD) and Vigileo (SVVig) was evaluated in Bland-Altman plot and trending in four-quadrant and polar plots. Bias ± limits of agreement (LOA) between SVOD and SVVig was 24 ± 37 ml (percentage error 45%). Concordance of the two methods from before to after a fluid load was 100%. Angular bias ± LOA was 12° ± 28°. Absolute values of SVOD and SVVig agreed poorly, but changes were highly concordant during fluid loads in aortic stenosis patients. The angular agreement indicated acceptable trending. The two measurement methods are not interchangeable in patients with aortic stenosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.