Abstract

BackgroundMarkerless systems to capture body motion require no markers to be attached to the body, thereby improving clinical feasibility and testing time. However, the lack of markers might affect the accuracy of measurements.ObjectiveThis study aimed to determine the absolute reliability and concurrent validity of the Kinect system with MotionMetrix software for spatiotemporal variables during running at a comfortable velocity, by comparing data between the combination system and two widely used systems—OptoGait and high-speed video analysis at 1000 Hz.MethodsIn total, 25 runners followed a running protocol on a treadmill at a speed of 12 km/h. The Kinect+MotionMetrix combination measured spatiotemporal parameters during running (ie, contact time, flight time, step frequency, and step length), which were compared to those obtained from two reference systems.ResultsRegardless of the system, flight time had the highest coefficients of variation (OptoGait: 16.4%; video analysis: 17.3%; Kinect+MotionMetrix: 23.2%). The rest of the coefficients of variation reported were lower than 8.1%. Correlation analysis showed very high correlations (r>0.8; P<.001) and almost perfect associations (intraclass correlation coefficient>0.81) between systems for all the spatiotemporal parameters except contact time, which had lower values. Bland-Altman plots revealed smaller systematic biases and random errors for step frequency and step length and larger systematic biases and random errors for temporal parameters with the Kinect+MotionMetrix system as compared to OptoGait (difference: contact time +3.0%, flight time −7.9%) and high-speed video analysis at 1000 Hz (difference: contact time +4.2%, flight time −11.3%). Accordingly, heteroscedasticity was found between systems for temporal parameters (r2>0.1).ConclusionsThe results indicate that the Kinect+MotionMetrix combination slightly overestimates contact time and strongly underestimates flight time as compared to the OptoGait system and high-speed video analysis at 1000 Hz. However, it is a valid tool for measuring step frequency and step length when compared to reference systems. Future studies should determine the reliability of this system for determining temporal parameters.

Highlights

  • MethodsThe use of marker-based motion capture technology has increased significantly in both research and diagnostics

  • The results indicate that the Kinect+MotionMetrix combination slightly overestimates contact time and strongly underestimates flight time as compared to the OptoGait system and high-speed video analysis at 1000 Hz

  • It is a valid tool for measuring step frequency and step length when compared to reference systems

Read more

Summary

Introduction

The use of marker-based motion capture technology has increased significantly in both research and diagnostics. Markerless systems do not require attaching any markers or sensors to the body, which substantially improves clinical feasibility and testing time. Studies focused on analyzing the validity of these systems under different circumstances are especially important In this context, a markerless motion capture system (ie, the Kinect) has received much attention from clinicians, sports practitioners, and researchers [1,3,4,5,6,7,8,9]. Markerless systems to capture body motion require no markers to be attached to the body, thereby improving clinical feasibility and testing time.

Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.