Abstract

ABSTRACT Purpose: To evaluate the agreement of different biometric parameters obtained using intraoperative spectral-domain optical coherence tomography (SD-OCT) and two swept-source optical coherence tomography (SS-OCT) based biometers. Methods: 102 eyes were assessed using the intraoperative SD-OCT integrated into the Catalys femtosecond-laser, and the IOLMaster 700 and Anterion SS-OCT-based-biometers. Central corneal thickness (CCT), anterior chamber depth (ACD), white-to-white (WTW), and lens thickness (LT) were measured. Results: There were statistically significant differences for CCT, ACD, WTW and LT between devices (p < 0.001). The mean difference for ACD ranged from −0.067 to −0.250 mm, with the largest mean difference being between the IOLMaster 700 and Catalys. CCT mean differences ranged from 7 to 32 µm, with the largest mean difference being between the Anterion and Catalys. For WTW, the comparison between the IOLMaster 700 vs Catalys showed the largest mean difference (0.38 mm). However, the mean differences for LT from all three devices were quite similar, ranging from −0.02 to −0.08 mm. Conclusions: SS-OCT biometers showed good agreement for ACD, CCT, WTW and LT. The SD-OCT showed ACD, CCT and WTW values that do not seem to be interchangeable with the SS-OCT biometers; however, this device did show excellent agreement in the case of LT.