Abstract
Purpose: To compare central corneal thickness (CCT) values obtained by Lenstar (LE), Pentacam (PC), specular microscopy (SM) and ultrasound pachymetry (UP) in healthy corneas and study their influence on intraocular pressure (IOP) readings determined by Goldmann applanation tonometry (GAT). Methods: CCT was measured in 76 healthy subjects by LE, PC, SM and UP. We established Lin's concordance correlation coefficient (ρ-C) between different techniques. The influence of CCT on GAT was established through univariate linear regression models, IOP being the dependent variable. Results: The highest ρ-C was found between LE and SM at 0.94 (95% CI: 0.91-0.96) and between LE and UP at 0.95 (95% CI: 0.94-0.97). IOP readings showed less variability when CCT was determined using LE (7.7%, B = 0.16; 95% CI: 0.004-0.28). Conclusions: Although CCT values obtained with UP, PC, SM and LE show good correlation, these methods are not completely interchangeable. The amount of IOP variation differs when CCT is determined using LE or SM.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.