Abstract
MR elastography (MRE) typically requires manual ROI placement to generate liver shear stiffness measurements. Among 419 patients (primarily children and young adults), a computer-based automated MRE processing tool and clinically reported manual ROI-based measurements generated similar results (mean bias = 0.13 kPa). The intraclass correlation coefficient was 0.94 and was at least 0.90 across common indications in male and female patients and in patients with and without elevated liver fat fraction. Automated analysis may promote postprocessing standardization and decrease reporting variability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.