Abstract

A long-term aging test of a Resistive Plate Chamber (RPC) was carried out with an intense gamma 137 Cs source. The detector was operated in avalanche mode and had the bakelite surface treated with linseed oil. After the irradiation the estimated dose, charge and fluence were approximately equal to the expected values after 10 years of operation in the CMS barrel region. During and after the irradiation, the RPC performance was monitored with cosmic muons and showed no relevant aging effects. Moreover, no variation of the bakelite resistance was observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.