Abstract

The wear performance of a polyurethane compliant layer (CL) material formed into an acetabular component and implanted into a sheep model of cemented total hip arthroplasty was assessed at 6, 12, 24, and 48 months. Four (11%) of 36 acetabular components debonded from the cement and one component was slightly loose at the cement-bone interface. There was no macroscopic evidence of fracture, wear, or deformation of the CL material on the articular surface of the acetabular components. Small numbers of polymeric wear particles was found in the hip synovial tissues of 10 sheep, most commonly in the early time groups, and were likely associated with initial wear of surface asperities. The wear performance of the CL was unchanged during a 48-month implantation period.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.