Abstract
This study questions the aging of non-implanted breast prostheses for a period of 9–60 months. Every 6 months, two non-implanted Natrelle™ prostheses were tested to measure the strength at break, the elongation at break, and the thickness of the shell. Then, the breaking stress was calculated from the preceding quantities. All these quantities were observed by separating the samples taken from the anterior and posterior sides of the prostheses. One-way ANOVA analyses (analysis of variance) were performed to define the influence of aging duration, lot membership, and side. In addition, the elongation at break and the thickness of the shell showed significant variations as a function of aging regardless of the side but without any trend emerging. For other quantities, there were significant disparities between the anterior and posterior sides of the prostheses, differences between prostheses from different lots, and similarities between prostheses from the same lot. Finally, the thickness is an important parameter. Since manufacturing is a manual process, it is necessary to check the thickness, which must be homogeneous on both sides. Always weaker on the anterior side than on the posterior side, it influences the mechanical properties. We recommend, like other studies, that its control be part of the quality controls during manufacturing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Mechanical Behavior of Biomedical Materials
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.