Abstract
Quantitative scanning micropipette contact method measurements are subject to the deleterious effects of reference electrode interference. The commonly used Ag/AgCl wire quasi-reference counter electrode in the miniaturized electrochemical cell of the scanning micropipette contact method was found to leak Ag+ into the electrolyte solution. The reduction of these Ag+ species at the working electrode surface generates a faradaic current, which significantly affects the low magnitude currents inherently measured in the scanning micropipette contact method. We demonstrate that, during the microscopic corrosion investigation of the AA7075-T73 alloy using the oil-immersed scanning micropipette contact method, the cathodic current was increased by the Ag+ reduction, resulting in positive shifts of corrosion potentials. The use of a leak-free Ag/AgCl electrode or an extended distance between the Ag/AgCl wire and micropipette tip droplet eliminated the Ag+ contamination, making it possible to measure accurate corrosion potentials during the oil-immersed scanning micropipette contact method measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.