Abstract

To compare the number of phosphatidylserine (PS)-exposing platelets obtained using the dual-platform approach and bead-based flow cytometry. Platelets were enumerated using the ADVIA 2010i instrument (Siemens AG). The numbers and percentages of PS-exposing platelets in 175 platelet products were determined using a FACSCalibur flow cytometer (Becton, Dickinson and Company) and counting beads. Our results showed good correlation (r2 = 0.96; P <.001) between the PS-exposing platelets obtained using counting beads and the dual-platform approach. The results of Bland-Altman analysis showed a bias of +46,449 cells per µL and a limit of agreement (LOA) from -197,863 to 290,762 cells per µL. Also, 8 measurements (5.0%) revealed a number of PS-exposing platelets outside the LOA ranges. Further, 21 measurements (12.0%) revealed greater than 2-fold changes in the number of PS-exposing platelets. The results suggest that the dual-platform approach is affordable and reliable for quantitating PS-exposing platelets as part of monitoring the quality of platelet products.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.