Abstract

A new technique has been developed to measure the size distribution and charge state of highly charged aerosols using an Electrical Low Pressure Impactor (ELPI). The internal charger was switched alternately on and off and the time between stable charge states found to be ∼ 10 s. The size distribution of aerosols was found when the charger was on, from which the charge distribution can be estimated when the charger is off using the current at each impactor stage. This method was tested in background conditions, when a candle was burning and when a negative air ioniser was used. The ELPI electrometers were not sensitive enough to accurately measure the charge state on background and candle air, but gave a value for air charged by an ioniser. Comparing results from the ELPI with other techniques showed inaccuracies in this method that need to be addressed before further use of this technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.