BackgroundIn a Fontan candidate, univentricular pacing may cause delay in interventricular conduction, which induces asynchronous contraction. Cardiac resynchronization therapy is expected to be an effective mode of therapy in such a case.Case presentationA 7-month-old girl, diagnosed with dextrocardia, congenitally corrected transposition of the great artery [situs solitus, L-loop, and L-transposition], ventricular septal defect, infundibular and pulmonary valvular stenosis, and straddling of the tricuspid valve, was considered as a candidate for the Fontan procedure. She had undergone Blalock-Taussig shunt, and epicardial univentricular pacemaker implantation for persistent complete atrioventricular block. She underwent a bidirectional cavopulmonary shunt concomitant with ventricular lead translocation from the morphological left ventricle to the morphological right ventricle. After discharge, ventricular dyssynchrony was noted and cardiac failure persisted. She was converted to cardiac resynchronization therapy (CRT) at 13 months of age. Two-dimensional speckle tracking imaging was used by cardiologists to determine the most suitable pacing site. CRT rapidly corrected the heart failure; thus, she underwent the Fontan procedure after 1.5 years. Five years have passed since the cardiac resynchronization therapy; her interventricular synchrony is maintained well and the level of brain natriuretic peptide remains within normal range.ConclusionWe describe the successful conversion from single ventricular pacing to CRT, in a case of congenitally corrected transposition of the great artery indicated for the Fontan procedure. The long-term prognosis of cardiac resynchronization therapy is undetermined in the pediatric population; therefore, further follow-up is required.
Highlights
In a Fontan candidate, univentricular pacing may cause delay in interventricular conduction, which induces asynchronous contraction
We describe the successful conversion from single ventricular pacing to cardiac resynchronization therapy (CRT), in a case of congenitally corrected transposition of the great artery indicated for the Fontan procedure
Because of persistent complete atrioventricular block, which appeared at the induction of the anesthesia for Blalock-Taussig shunt (BTS), she had undergone an epicardial univentricular pacemaker implantation on the morphological left ventricle, at 5 months of age
Summary
We describe the successful conversion from single ventricular pacing to CRT, in a case of congenitally corrected transposition of the great artery indicated for the Fontan procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.