Abstract
ObjectiveTo identify adverse events (AEs) related to suction electrocautery use during adenotonsillectomy. MethodsThe US Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database was searched using the terms “suction cautery,” “suction electrocautery,” “suction Bovie,” and “suction coagulator” from January 2014 to December 2023. Results165 AE reports were gathered from the MAUDE database medical device reports (MDRs). 36 met inclusion criteria. Patient injuries were found in 22 (61.1 %) reports and device malfunction events were found in 14 (38.9 %) reports. All patient injuries were thermal burns (N = 22, 100 %). Location of burn injuries included the lip (N = 6, 27.3 %), oral commissure (N = 5, 22.7 %), and tongue (N = 4, 18.2 %). The most common cause of an AE was inadequate device insulation (N = 7, 19.4 %). ConclusionThe suction electrocautery apparatus may malfunction and cause patient burn injuries. Device failures mainly result from inadequate device insulation, coagulation problems, and detachment of device components. Surgeons must be aware of these potential complications and counsel parents and patients regarding AEs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: American Journal of Otolaryngology--Head and Neck Medicine and Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.