Abstract

A multichannel non-linear frequency compression algorithm was evaluated in comparison to conventional amplification hearing aids using a test of speech understanding in noise (Oldenburger Satztest-OLSA) and subjective questionnaires. The new algorithm compresses frequencies above a pre-calculated cut off frequency and shifts them to a lower frequency range, thereby providing high-frequency audibility. Low-frequencies, below the compression cut off frequency, are amplified normally. This algorithm is called SoundRecover (SR). In this study, 11 experienced hearing aid users with a severe to profound sensorineural hearing loss were tested. Seven subjects showed enhanced levels of understanding in noise (OLSA) using frequency compression. However, 4 out of the 11 subjects could not benefit from the high-frequency gain. Evaluation using questionnaires demonstrated an increased level of satisfaction after 2 months of experimental devices wearing (p = 0.08) and after 4 months of wearing (p = 0.09), respectively, compared to conventional hearing instruments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.