Laser ablation molecular isotopic spectrometry (LAMIS) recently has been investigated for analysis of the boron isotopic composition in ambient air at atmospheric pressure. The initial precision of 100–400‰ (2σ) for the 10B/11B isotopic ratio was less than results obtained from other analytical methods like TIMS or ICP–MS. This paper describes how accuracy and precision for boron isotopic ratio using LAMIS can be improved to the few per mil level. Several optimization procedures, viz., spectral region of analysis, effect of flicker noise from matrix, spectral normalization and pre-treatment procedures, were studied. This paper reports a precision of 9‰ (2σ) for the 10B/11B ratio using alternative spectral normalization and pre-treatment procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.