Abstract

BMS-986142 is a Bruton's tyrosine kinase inhibitor under development to treat several disease types. The compound contains three chiral elements: one chiral center and two chiral axes, resulting in three potential atropisomeric impurities in its drug substance and drug products. Separation of BMS-986142 atropisomers has been successfully achieved on an achiral polar-embedded C18 column in reversed-phase liquid chromatography (RPLC) and on polysaccharide-based chiral columns in RPLC and supercritical fluid chromatography (SFC). Compared to the RPLC chiral separation, the SFC atropisomeric separation on a sub-2 μm immobilized cellulose-based column is much more efficient and environmentally friendly. The analysis time in SFC was reduced by 8-fold compared to that in RPLC, and the method sensitivity in SFC on the sub-2 μm chiral column in 3.0mm I.D. was 2 to 4-fold better than that on 3 μm chiral columns in 4.6mm I.D.. Furthermore, our study suggests that the contribution to band broadening from the extra column volume (ECV) of modern commercial SFC instrument was not negligible for a 3.0mm I.D.×100mm column packed with 1.6 μm particles. This result reaffirms that there is a great need for further improvement of SFC instrument design in order to realize the full theoretical efficiency of both sub-2 μm achiral and chiral columns.