Abstract

DNA electrophoresis in gels and solutions of agarose and polyacrylamide was objectively evaluated with regard to separation efficiency at optimal polymer concentrations. In application to DNA fragments, polyacrylamide gels were superior for separating fragments of less than 7800 bp, and agarose gels are the best choice for larger fragments. Agarose solutions are nearly as good as polyacrylamide gels for small DNA (< 300 bp). Agarose solutions have a higher efficiency than polyacrylamide solutions for DNA of less than 1200 bp. Separation efficiency sharply decreases with increasing length of DNA. Retardation in polyacrylamide solutions was found to depend on polymer length in a biphasic fashion. The choice of resolving polymer concentrations depends on the progressive stretching of DNA in proportion to polymer concentration. The rate of that stretching appears higher in polyacrylmide solution than in gels or in liquid or gelled agarose. Application of polymer solutions to capillary electrophoresis raises further problems concerning agarose plugs, DNA interactions with the polymers, operation at low field strength and long durations as well as detection sensitivity.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call