Abstract

Oxinium™ (Smith & Nephew, Memphis, TN, US) has been used in hip arthroplasty since 2003. The surface coating is hard and provides low wear rates but if this surface coating is damaged, the soft metal core is at risk of accelerated wear. Previous reports have described accelerated wear following intra and postoperative hip dislocation. We report a case of advanced wear of an in situ Oxinium™ femoral head implant following a cracked acetabular liner. The liner had disengaged from the titanium shell, allowing the Oxinium™ head to articulate directly with the shell. The disengaged liner led to dislocation of the Oxinium™ head, with associated pronounced wear of the head and the acetabular cup. The patient had a successful revision procedure. We advise close follow-up of patients with Oxinium™ implants, especially if associated with dislocation and closed reduction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.