Abstract
A new β-ray-induced X-ray measurement system equipped with a silicon drift detector, which was named “Advanced-BIXS”, was constructed to study in detail retention behavior of surface tritium by measurements of low energy X-rays below 1keV such as C(Kα) and O(Kα) as well as high energy X-rays induced by β-rays from tritium. In this study, basic performance of the present system has been examined using various tritium-containing samples. It was seen that energy linearity, energy resolution and sensitivity were quite enough for measurements of low energy X-rays induced by β-rays. Intensity of characteristic X-rays emitted from the surface and/or bulk of a tritium-containing sample was lowered by argon used as a working gas of the Advanced-BIXS. Pressure dependence of transmittance of C(Kα) and Fe(Kα) was examined as examples of low and high energy X-rays, and it was able to represent by using the mass absorption coefficient in argon. It was concluded, therefore, that the present system has high potentiality for nondestructive measurements of tritium retained in surface layers and/or bulk of fusion related materials.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.