Abstract
Primary stability of artificial acetabular cups, used for total hip arthroplasty, is required for the subsequent osteointegration and good long-term clinical results of the implant. Although closed-cell polymer foams represent an adequate bone substitute in experimental studies investigating primary stability, correct numerical modelling of this material depends on the parameter selection. Material parameters necessary for crushable foam plasticity behaviour were originated from numerical simulations matched with experimental tests of the polymethacrylimide raw material. Experimental primary stability tests of acetabular press-fit cups consisting of static shell assembly with consecutively pull-out and lever-out testing were subsequently simulated using finite element analysis. Identified and optimised parameters allowed the accurate numerical reproduction of the raw material tests. Correlation between experimental tests and the numerical simulation of primary implant stability depended on the value of interference fit. However, the validated material model provides the opportunity for subsequent parametric numerical studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Computer Methods in Biomechanics and Biomedical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.