Calculation of the geometric part G of the profile shape is part of the recently developed fundamental parameters profile model. A Monte Carlo raytracing algorithm is used to describe realistic conditions such as multiple slits/collimators, dimension of secondary monochromator and complicated specimen shape. To model the X-ray source more exactly, we successfully introduced a correction for X-ray tube tails. A relevant part (6%…30%) of the X-rays is produced not from a proper rectangular line focus shape but from supplementary tails on both sides of the optical focus. Thereby, the measured profile shapes differ from those simply derived from a box-shaped focus. Applying an X-ray tube tails correction allows the use of fundamental parameter description for line profile (micro strain) analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.