Abstract
The optimum biprism position as suggested by Lichte (Ultramicroscopy 64 (1996) 79 [10]) was implemented into a state-of-the-art transmission electron microscope. For a setup optimized for atomic resolution holograms with a width of 30nm and a fringe spacing of 30pm, we investigated the practical improvements on hologram quality. The setup is additionally supplemented by a second biprism as suggested by Harada et al. (Applied Physics Letters 84 (2004) 3229 [12]). In order to estimate the possibilities and limitations of the double biprism setup, geometric optics arguments lead to calculation of the exploitable shadow width, necessary for strong reduction of biprism-induced artefacts. Additionally, we used the double biprism setup to estimate the biprism vibration, yielding the most stable imaging conditions with lowest overall fringe contrast damping. Electron holograms of GaN demonstrate the good match between experiment and simulation, also as a consequence of the improved stability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.