Abstract
In this paper, we present an advanced triggering method for digital oscilloscopes for measurements involving complex waveforms that are challenging even today. Based on the associative-memory technique and on utilizing a new weighted-hamming-distance parameter, this technique is capable of providing stable oscilloscope triggering, even for complex waveforms where the commonly used voltage level triggering method fails to perform adequately. The proposed technique has been verified by exhaustive software simulations using Matlab and Xilinx foundation series softwares. Furthermore, the technique has been implemented and validated on a XCV300EPQ240 field-programmable gate array device and has given successful initial results using several different kinds of test waveforms
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.