Abstract

A highly selective and sensitive adsorptive stripping voltammetric procedure for trace measurements of thallium in presence of o-cresolphthalein (OCP) is described. The response is not affected by the presence of excess tin, lead or indium, which can interfere severely in analogous anodic stripping measurements. The adsorptive strategy also offers enhanced sensitivity over anodic stripping measurements. The detection limit is 60 ng/1 (3 × 10 −10 M) with 2-min accumulation, and the relative standard deviation (at the 5 υg/1 level) is 3.6. Optimal conditions were found to be 0.01 M sodium hydroxide containing 5 × 10 −7 M OCP, a preconcentration potential of −0.30 V and a differential pulse waveform.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.