Abstract
This article presents for the first time a new working electrode with a long service life- the bismuth-plated array of carbon composite microelectrodes for the simple, fast and sensitive determination of quercetin by adsorptive stripping voltammetry. The main experimental conditions were selected. The calibration graph was linear from 1 × 10-9 to 2 × 10-8 mol L-1 with an accumulation time of 60 s. The detection limit was equal to 4.8 × 10-10 mol L-1. The relative standard deviation for 2 × 10-8 mol L-1 of quercetin was 4.4% (n = 7). Possible interference effects resulting from the presence of other organic and surface active compounds and interfering ions were studied. The developed procedure was successfully applied to determine quercetin in pharmaceutical preparations and the spiked urine samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.