Abstract
The electrochemical behaviour of the antidepressant drug sulpiride (SP) at a hanging mercury drop electrode (HMDE) is investigated. Linear sweep cathodic stripping voltammetry (LSCSV) was used to determine sulpiride in the presence of 0.01 M sodium acetate medium pH 10.5 and 25±1°C. Different parameters such as, supporting electrolyte, pH, accumulation potential, scan rate, accumulation time and ionic strength, were tested to optimize the conditions for the determination of SP. The adsorbed form is reduced irreversibly. The linear concentration range is from 2×10 −9 to 5×10 −8 M SP. Experimentally, 2×10 −9 M (0.68 ppb) with accumulation time 60 s can be determined successfully. Furthermore, a theoretical detection limit of 2×10 −10 M (0.068 ppb) Sp was calculated. The interferences of some metal ions, ascorbic acid and some amino acids were studied. The method was applied to the analysis of tablets and spiked urine, with recoveries of 104±3 and 101±3, and the relative standard deviation of 3.3 and 3.4%, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.