Abstract
Two sensitive voltammetric techniques (adsorptive and catalytic cathodic stripping) are described for the determination of timonacic, in the presence of copper or nickel ions, respectively. The measured peaks which correspond to the reduction or copper (I) or nickel – timonacic complexes are measured at -773 and -782 mV, respectively. The different experimental parameters have been carefully studied. The methods have been fully validated. The detection limits were 1.5 and 13.6 ng mL -1 , respectively. The methods have been applied for the determination of timonacic in pharmaceutical tablets. The obtained results have been compared statistically with those obtained from a published method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.