Abstract
Uranium (VI) can be determined by adsorptive voltammetric techniques, as its chloranilic acid complex, over a wide concentration range. Differential pulse polarography is useful for quantification of uranium between 0.1 and 1.5 mg/l; for the range from 10 to 500 μg/l differential pulse voltammetry and for ultra-trace analysis between 0.024 and 40 μg/l adsorptive stripping voltammetry are preferred. The standard deviation for the 3σ-detection limit of 24 ng/l uranium was found to be 8%. In the trace analysis of metals in aquatic environmental systems by adsorptive stripping voltammetry it is normally necessary to decompose polluted water samples by UV irradiation or microwave digestion. The advantage of the developed method is the fact that no sample pretreatment is necessary.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.