Abstract
Sinotubular junction size in aortic valve reimplantation procedures is usually predetermined on the basis of mathematical calculations and intraoperative measurements. We propose a new method for aortic valve reimplantation by which intraoperative measurements can be eliminated and sinotubular junction size adjusted after cross clamp removal to fit the patient's need. Aortic valve commissures are reimplanted in the expandable skirt of a Valsalva (Vascutek, Renfrewshire, Scotland) graft to realize an oversized sinotubular junction that is subsequently reduced to the proper size by wrapping, with Dacron rings of decreasing size, the neo-sinotubular ridge under transesophageal echocardiographic guidance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.