A novel method for simplifying adduct patterns to improve the detection and identification of peptide toxins using matrix-assisted laser desorption/ionization (MALDI) time-of-flight (TOF) mass spectrometry is presented. Addition of 200 microM zinc sulfate heptahydrate (ZnSO(4) . 7H(2)O) to samples prior to spotting on the target enhances detection of the protonated molecule while suppressing competing adducts. This produces a highly simplified spectrum with the potential to enhance quantitative analysis, particularly for complex samples. The resulting improvement in total signal strength and reduction in the coefficient of variation (from 31.1% to 5.2% for microcystin-LR) further enhance the potential for sensitive and accurate quantitation. Other potential additives tested, including 18-crown-6 ether, alkali metal salts (lithium chloride, sodium chloride, potassium chloride), and other transition metal salts (silver chloride, silver nitrate, copper(II) nitrate, copper(II) sulfate, zinc acetate), were unable to achieve comparable results. Application of this technique to the analysis of several microcystins, potent peptide hepatotoxins from cyanobacteria, is illustrated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.