Abstract

Severe bone atrophy jeopardizes the success of endosseous implants. This technical note aims to present the innovative concept of additively manufactured sub-periosteal jaw implants (AMSJIs). Digital datasets of the patient’s jaws and wax trial in occlusion are used to segment the bone and dental arches, for the design of a sub-periosteal frame and abutments in the optimal location related to the dental arch and for the design of the suprastructure. The implants and suprastructure are three-dimensionally (3D) printed in titanium alloy. The provisional denture is 3D-printed in polymer. AMSJIs offer an alternative approach for patients with extreme jaw bone atrophy. This report refers to the use of this technique for full maxillary rehabilitation, but partial defects in either jaw and extended post-resection defects may also be approached using the same technique. This customized, prosthesis-driven reverse-engineering approach avoids bone grafting and provides immediate functional restoration with one surgical session.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.