Abstract

Hydrophobic interaction chromatography (HIC) is a traditional technique used for the separation, purification, and characterization of proteins. As the number of antibody-drug conjugates (ADCs) continues to increase in clinical trials, HIC and other orthogonal methods utilizing changes in hydrophobicity are being used for ADC characterization and analysis. Unlike other techniques, HIC uniquely allows for protein analysis under mild nondenaturing conditions that preserve the native structure and activity of the molecules. Analysis of the ADC in its native form is advantageous. Herein, we describe a generic HIC protocol for the screening, analysis, and characterization of ADCs using an ammonium sulfate buffer and a high-pressure liquid chromatography system. Parameters affecting data quality and interpretation are addressed. In addition, several recommendations are included for method optimization and troubleshooting.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call