Abstract

An adaptive algorithm (CAMAC, control advance moving average controller) to control multiinput/multioutput physiological systems has been implemented and tested. The algorithm is a self-tuning controller that determines the input on the basis of the expected difference between the output and desired output at a time interval equal to or greater than the system dead time. The algorithm was used to simultaneously control mean arterial pressure and cardiac output in anesthetized dogs by the simultaneous computer-controlled infusion of sodium nitroprusside and dobutamine. The results demonstrate the feasibility of using CAMAC for multivariable drug delivery, but they indicate the need for further work before clinical applications are attempted.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.