Abstract
Abstract We report successful operation of the first X-ray monochromator with built-in correction for the thermal strain field created by a high power X-ray beam. The 27 pole wiggler magnet at the National Synchrotron Light Source provides a total power load of 330 W in a beam 60 mm wide. A (220) silicon monochromator system has been tested; the compensated deformation field contributes only 2 arc sec full width at half maximum intensity to the monochromator Darwin width of high order Bragg reflections.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.