Abstract

Severe valvular regurgitation can lead to pulmonary hypertension, atrial fibrillation and heart failure. Vena contracta width is used to estimate the severity of the regurgitation. Parameters affecting visualization of color Doppler have a significant impact on the measurement. We propose a data-driven method for automated adjustment of color gain based on the peak power of the color Doppler signal in the vicinity of the vena contracta. A linear regression model trained on the peak power was used to predict the orifice diameter. According to our study, the color gain should be set to about 6 dB above where color Doppler data completely disappears from the image. Based on our method, orifices with reference diameters of 4, 6.5 and 8.5 mm were estimated with relative diameter errors within 18%, 12% and 14%, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.