Abstract

We have developed a new adaptive clutter rejection technique where an optimum clutter filter is dynamically selected according to the varying clutter characteristics in ultrasound color Doppler imaging. The selection criteria have been established based on the underlying clutter characteristics (i.e., the maximum instantaneous clutter velocity and the clutter power) and the properties of various candidate clutter filters (e.g., projection-initialized infinite impulse response and polynomial regression). We obtained an average improvement of 3.97 dB and 3.27 dB in flow signal-to-clutter-ratio (SCR) compared to the conventional and down-mixing methods, respectively. These preliminary results indicate that the proposed adaptive clutter rejection method could improve the sensitivity and accuracy in flow velocity estimation for ultrasound color Doppler imaging. For a 192 x 256 color Doppler image with an ensemble size of 10, the proposed method takes only 57.2 ms, which is less than the acquisition time. Thus, the proposed method could be implemented in modern ultrasound systems, while providing improved clutter rejection and more accurate velocity estimation in real time.