Abstract
Balanced detection optical coherence tomography (BD-OCT) enables near-shot noise-limited imaging by suppressing wavelength-dependent relative intensity noise (RIN) originating from the light source. In spectral-domain BD-OCT (SD-BD-OCT), the level of RIN suppression relies on the co-registration accuracy of the spectra simultaneously captured by two independent spectrometers. However, existing matching methods require careful pre-calibration using a RIN-dominated dataset or subjective post-processing using a signal-dominated dataset. We developed an adaptive subpixel matching approach, referred to as adaptive balance, that can be applied to any SD-BD-OCT dataset regardless of RIN or signal level without the need for pre-calibration. We showed that adaptive balance performed comparable to or better than reported methods by imaging phantoms with varying spectrometer camera gain, exposure time, and supercontinuum laser repetition rate. We further demonstrated the benefits of adaptive balance in human retinal imaging.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.