Abstract

Graded-index (GRIN) lenses serve as a key component for miniature endoscopes because of their small diameters and ease of assembly. However, the nonaplanatic nature of GRIN lenses causes inherent spatial aberrations that lower image resolution and sharpness. Here we present the diagnosis of the aberrations in GRIN probes and the use of adaptive optics to compensate for the wavefront errors in the endoscope. Two different operation schemes based on preset and in situ measurements are demonstrated, both resulting in a substantial reduction of the wavefront error from 0.42 to <0.1 μm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.