Abstract

Adaptive optics is implemented in a confocal scanning fluorescence microscopy using a wavefront sensorless correction scheme. Using the image sharpness as the optimization metric, aberration correction is performed to compensate both system- and specimen-induced aberrations by using stochastic parallel gradient descent algorithm based upon Zernike polynomial modes. We demonstrate the idea of using phantom fluorescence samples experimentally. Enhanced imaging contrast and improved signal level are achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.