Abstract
We demonstrate aberration correction in two-photon microscopy. Specimen-induced aberrations were measured with a modal wavefront sensor, implemented using a ferro-electric liquid crystal spatial light modulator (FLCSLM). Wavefront correction was performed using the same FLCSLM. Axial scanned (xz) images of fluorescently labelled polystyrene beads using an oil immersion lens show restored sectioning ability at a depth of 28 &mgr;m in an aqueous specimen.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.