Abstract

With the application of a heater element to a cryostat of a low-temperature Guinier diffractometer and camera a working range from 12 to 700 K is achieved. The heating/cooling rates in the low-temperature range are from 0.5 to 160 K/h and in the high-temperature range from 0.5 to 600 K/h. The sample chamber can be either evacuated or filled with a gas up to a pressure of 0.2 MPa. To increase the accuracy of x-ray intensity measurements rotation of the sample is possible within the whole temperature range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.