Abstract

We have developed a quantitative PCR assay (LightCycler* using the pair of primers JW58 and JW59 for the detection of the 35-fold repeated B gene of oxoplasma gondii. This real-time PCR, using fluorescence resonance energy transfert (FRET) hybridization probes, allows the quantification of . gondii with several technical requirements not previously described: i) an internal amplification control (co-amplified in a single tube with the same primers), ii) Uracil-N-Glycosylase and iii) a standard curve corresponding to a serial dilution from a calibrated suspension of T. gondii ranging from 40 to 4.106( )parasites in one ml of amniotic fluid (1 to 105( ) . gondii/PCR). In artificial samples, one parasite could be detected if at least three reactions were performed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.