Abstract

BackgroundRemoving a head during isolated acetabular revision surgery can cause damage to the stem taper surface from extraction tool contact. Implanting a ceramic head on the damaged stem taper might elevate the fracture risk, which can be mitigated with the use of titanium adapter sleeves. The aim of this study was to investigate whether the improved fracture strength of modern generation ceramic heads allows the direct implantation on damaged stem tapers without an adapter sleeve. MethodsFinite element models of taper junctions with and without adapter sleeve were generated. Different stem taper damages were modelled to investigate the influence on the ceramic head fracture load under axial compression. FindingsHeads without adapter sleeves exhibited slightly higher or equal fracture strengths compared with sleeved heads for most scenarios. However, a small metal elevation on the stem taper caused a drastic decrease of the fracture strength if no adapter sleeve was used (−96%). The sleeved head was not influenced by the metal elevation damage. InterpretationAdapter sleeves are essential to ensure patient safety and prosthesis longevity whenever implanting ceramic heads on used stem tapers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.