Abstract

A proton transfer reaction mass spectrometer (PTR-MS) instrument was adapted to employ NO+ as a chemical reagent ion without any hardware changes by switching the reagent ion source gas from water vapor to dry air. Ionization of dry air within the hollow cathode ion source generates a very intense source of NO+ with only a minor impurity of NO2+. The intensities of the primary NO+ reagent ion and the unwanted impurity NO2+ are controllable and dependent on the operational conditions of the hollow cathode ion source. Ion source tuning parameters are described, which maintain an intense source of NO+ while keeping the impurity NO2+ signal to less than 2% of the total reagent ion intensity. This method is applied to the detection of 1,3-butadiene. NO+ reacts efficiently with 1,3-butadiene via a charge exchange reaction to produce only the molecular ion, which is detected at m/z 54. Detection sensitivities of the order of 45 pptv for a 1-s measurement of 1,3-butadiene are demonstrated. We present the first real-time on-line sub parts per billion measurement of 1,3-butadiene in the ambient atmosphere. The only likely interference is from 1,2-butadiene. Concurrent measurements of benzene are provided and suggest that the vehicular emissions are the predominant source of 1,3-butadiene in a suburban Boston area monitoring location.