Abstract

IntroductionFemoral head-neck modularity in total hip arthroplasty (THA) is advantageous but taper corrosion at the trunnion can result in implant failure. We report two cases of acute catastrophic trunnion failure with a TMZF alloy cementless stem. MethodsDemographic, clinical, radiographic and operative data including implant retrieval was recorded and is presented. ResultsCase 1: A 79 year old farmer presented with sudden onset of hip pain and an inability to weight bear. He underwent a cementless large diameter stemmed metal-on-metal system (MITCH acetabular component, 56mm cobalt chrome head 4.5 lateralised Accolade TMZF, Stryker) nine years previously. He denied symptoms prior to his presentation to the Emergency Department.Case 2: An 86 year old gentleman presented with sudden onset of hip pain and inability to weight bear. He underwent a cementless large diameter stemmed metal-on-poly THA (Trident acetabular component, X3 polyethylene insert, 44mm cobalt chrome head, 4.5 lateralised Accolade TMZF, Stryker)nine years previously. This man had been complaining of mild hip symptoms prior to presentation. ConclusionPatients that have received TMZF alloy cementless stems coupled with CoCr alloy heads are at risk of catastrophic trunnion failure. Importantly, background trunnion corrosion may occur silently and present emergently irrespective of surveillance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.