Abstract

A 58-year-old male who underwent cataract extraction with combined intraocular lens and Hydrus® Microstent (Ivantis Inc, Irvine, CA, US) implantation 2 years ago in the right eye (OD) due to advanced glaucoma presented with blurry vision in right eye (OD) for 3 months. The visual acuity was 20/60 and slit-lamp examination indicated mild anterior chamber inflammation with unexposed, functioning tube shunt superotemporally in OD. Optical coherence tomography demonstrated cystoid macular edema (CME) with subretinal fluid. Fluorescein angiography demonstrated petaloid pattern leakage of CME. Gonioscopy revealed a kinked appearance of a Hydrus® Microstent protruding into the anterior chamber and causing iris chafing. Topical ketorolac tromethamine and prednisolone acetate were started. At the 2nd month of follow-up, the anterior chamber was quiet, and the CME resolved completely. Protruded kinked Hydrus® Microstent may lead to acute iridocyclitis and CME through iris chafing, which may be responsive to topical anti-inflammatory drops.